Acta Vet. Brno 2012, 81: 319-323

https://doi.org/10.2754/avb201281030319

Determination of fluoride in spices using microwave induced oxygen combustion

Emanuel Šucman, Jiří Bednář

University of Veterinary and Pharmaceutical Sciences Brno, Faculty of Veterinary Hygiene and Ecology, Institute of Biochemistry, Chemistry and Biophysics, Brno, Czech Republic

Fluorine is essential in human and/or animal nutrition; therefore, so it is important to know its concentration in a diet. One of the possible sources of fluorine for humans is spice, containing various amounts of this trace element. This work describes the method for fluoride determination in various kinds ofspices using microwave-supported sample preparation in high pressure oxygen atmosphere followed by potentiometry with a fluoride ion-selective electrode. Parameters of the microwave device for combustion procedure were checked and optimized in order to find settings ensuring complete sample combustion and/or absorption of the analyte in the absorption solution. For the ion-selective electrode measurement, the technique of standard straight line was chosen. Concentrations of fluorides in spices and spice blends under investigation ranged from 3.15 mg·kg-1 to 26.08 mg·kg-1. In order to check the accuracy of the method Certified Reference Material Fluoride in Vegetation NIST 2695 was used and a good agreement between certified and found values was found. The precision expressed as the relative standard deviation ranged from 0.6% to 5.0%. The method is fast, accurate and reliable for this kind of analysis. In recent literature data on fluoride concentrations in spices and/or spice blends have not been found.

References

11 live references